Wear of ceramic-on-carbon fiber-reinforced poly-ether ether ketone hip replacements

  • Claire L. Brockett
  • , Gemma John
  • , Sophie Williams
  • , Zhongmin Jin
  • , Graham H. Isaac
  • , John Fisher
  • University of Leeds
  • Johnson & Johnson

科研成果: 期刊稿件文章同行评审

76 引用 (Scopus)

摘要

Total hip replacement has been a successful surgical intervention for over 50 years, with the majority of bearings using a polyethylene cup. Long-term failure due to osteolysis and loosening has been widely documented and alternative bearings have been sought. A novel carbon fiber-reinforced poly-ether ether ketone (CFR-PEEK) cup was investigated through experimental friction and wear studies. Friction studies demonstrated the bearings operated in a boundary lubrication condition, with friction factors higher than those for other hip replacement bearings. The wear study was conducted with 36 mm diameter bearings tested against Biolox Delta heads for a period of 10 million cycles. The mean volumetric wear rate was 0.3 mm 3/Mc, indicating the ceramic-on-CFR-PEEK bearing to be a very low wearing option for total hip replacement.

源语言英语
页(从-至)1459-1465
页数7
期刊Journal of Biomedical Materials Research - Part B Applied Biomaterials
100 B
6
DOI
出版状态已出版 - 8月 2012