Validation of a Monte Carlo simulation of the Inveon PET scanner using GATE

  • Lijun Lu
  • , Houjin Zhang
  • , Zhaoying Bian
  • , Jianhua Ma
  • , Qiangjin Feng
  • , Wufan Chen
  • Southern Medical University

科研成果: 期刊稿件 › 文章 › 同行评审

10 引用 (Scopus)

摘要

The purpose of this study is to validate the application of GATE (Geant4 Application for Tomographic Emission) Monte Carlo simulation toolkit in order to model the performance characteristics of Siemens Inveon small animal PET system. The simulation results were validated against experimental/published data in accordance with the NEMA NU-4 2008 protocol for standardized evaluation of spatial resolution, sensitivity, scatter fraction (SF) and noise equivalent counting rate (NECR) of a preclinical PET system. An agreement of less than 18% was obtained between the radial, tangential and axial spatial resolutions of the simulated and experimental results. The simulated peak NECR of mouse-size phantom agreed with the experimental result, while for the rat-size phantom simulated value was higher than experimental result. The simulated and experimental SFs of mouse- and rat- size phantom both reached an agreement of less than 2%. It has been shown the feasibility of our GATE model to accurately simulate, within certain limits, all major performance characteristics of Inveon PET system.