跳到主要导航 跳到搜索 跳到主要内容

Ultrasound Transcranial Imaging Based on Fast Coherent-Time-Delay and Correlative Pixel-Based Beamforming

  • Chen Bai
  • , Xinyu Zhang
  • , Xiaoyang Qiao
  • , Yuchao Sang
  • , Hui Zhong
  • , Mingxi Wan
  • Xi'an Jiaotong University
  • CAS - Xi'an Institute of Optics and Precision Mechanics

科研成果: 书/报告/会议事项章节会议稿件同行评审

4 引用 (Scopus)

摘要

Adaptive beamforming, such as the fast minimum variance (FMV)algorithm, has been developed to improve the contrast-to-noise ratio (CNR)in plane wave imaging with extremely reduced computation. However, in ultrasound trans cranial imaging, FMV will be confronted with challenges raised by echoes with low intensities, especially when calculating the beamformer contributions at points which are in a distance from the beam center line. In order to improve the CNR and resolution in low echo beamforming during trans cranial imaging, this paper proposed an adaptive beamforming which combined with fast minimum variance (FMV)algorithm and unified pixel-based beamforming, i.e. fast coherent-time-delay and correlative pixel-based (FCCP)beamforming. By considering the influence of neighboring pixels during the beamforming, this method can remove undesirable information instead of increasing diagnostic information in trans cranial imaging. The influence of neighboring pixels was reflected by coherent-time-delay and coherent coefficients of each covariance matrix. After the simulation and phantom study, the results prove that CNR and the resolution are indeed ameliorated be FCCP, compared with delay-and-sum (DAS)and conventional minimum variance (MV)algorithm. In simulation, the CNR was augmented by 6 dB as well as the lateral resolution was improved by from 0.1536 mm from 0.1740 mm. FCCP is the combination of compound coherent-time-delay and correlative pixel-based, which could improves the resolution and contrast if image, and FMV. FMV is involved for the high complexity caused by matrix inversion, the computation was simplified by principal component analysis (PCA)when calculate the covariance matrix.

源语言英语
主期刊名2018 IEEE International Ultrasonics Symposium, IUS 2018
出版商IEEE Computer Society
ISBN(电子版)9781538634257
DOI
出版状态已出版 - 17 12月 2018
活动2018 IEEE International Ultrasonics Symposium, IUS 2018 - Kobe, 日本
期限: 22 10月 201825 10月 2018

出版系列

姓名IEEE International Ultrasonics Symposium, IUS
2018-October
ISSN(印刷版)1948-5719
ISSN(电子版)1948-5727

会议

会议2018 IEEE International Ultrasonics Symposium, IUS 2018
国家/地区日本
Kobe
时期22/10/1825/10/18

学术指纹

探究 'Ultrasound Transcranial Imaging Based on Fast Coherent-Time-Delay and Correlative Pixel-Based Beamforming' 的科研主题。它们共同构成独一无二的学术指纹。

引用此