Transient elastohydrodynamic lubrication analysis of ultra-high molecular weight polyethylene hip joint replacements

科研成果: 期刊稿件文章同行评审

32 引用 (Scopus)

摘要

The cyclic variation in both the load and speed experienced during walking was considered in an elastohydrodynamic lubrication (EHL) analysis for artificial hip joint replacements in this study. A general numerical procedure was developed to take both the entraining and squeeze-film actions into the solution of the Reynolds equation in the spherical ball-in-socket coordinate, simultaneously with the elasticity equation, using the Newton-Raphson method. The numerical procedure developed was then applied to an example of hip joint replacements employing an ultra-high molecular weight polyethylene (UHMWPE) acetabular cup against either a metallic or ceramic femoral head under simplified cyclic load and speed conditions. The predicted minimum film thickness was found to stay remarkably constant, despite a large change in the angular velocity and the load. This was attributed to the combined effect of entraining and squeeze-film actions in generating, replenishing and maintaining the lubricating film in artificial hip joint replacements. Furthermore, it was pointed out that the average transient minimum film thickness predicted throughout one cycle was very close to that under quasi-static conditions based upon the average angular velocity and load.

源语言英语
页(从-至)409-420
页数12
期刊Proceedings of the Institution of Mechanical Engineers, Part C: Journal of Mechanical Engineering Science
216
4
DOI
出版状态已出版 - 2002
已对外发布

学术指纹

探究 'Transient elastohydrodynamic lubrication analysis of ultra-high molecular weight polyethylene hip joint replacements' 的科研主题。它们共同构成独一无二的指纹。

引用此