跳到主要导航 跳到搜索 跳到主要内容

Time response of helium scintillation fission neutron detector

  • Xiaodong Zhang
  • , Mengtong Qiu
  • , Jianfu Zhang
  • , Xiaoping Ouyang
  • , Liang Chen
  • , Jinliang Liu
  • Northwest Institute of Nuclear Technology

科研成果: 期刊稿件 › 文章 › 同行评审

摘要

A new helium scintillation fission neutron detector was designed to detect the neutrons in the mixed pulse fission radiation field. The time response of the detector was investigated through theoretical and experimental methods. The flying time of the fragments and recoiled helium nucleus in helium scintillator were calculated based on the experimental formulae and method of Monte Carlo. The formula of detector's time response was deduced from the convolution method. The result of the calculating detector's wave indicated that the risetime was about 19.5 ns and FWHM was about 31.0 ns. The time response was measured on the dense plasma focus device (DPF) neutron generator. The theoretical results of time response were consistent with experimental results.

源语言英语
页(从-至)1807-1811
页数5
期刊Qiangjiguang Yu Lizishu/High Power Laser and Particle Beams
卷25
期7
DOI
出版状态已出版 - 7月 2013
已对外发布是

学术指纹

探究 'Time response of helium scintillation fission neutron detector' 的科研主题。它们共同构成独一无二的学术指纹。

引用此