TY - JOUR
T1 - Tibial post loading increases the risk of aseptic loosening of posterior-stabilized tibial prosthesis
AU - Chen, Zhenxian
AU - Han, Jianian
AU - Zhang, Jing
AU - Peng, Yinghu
AU - Guo, Lei
AU - Chen, Shibin
AU - Jin, Zhongmin
N1 - Publisher Copyright:
© IMechE 2024.
PY - 2024/8/1
Y1 - 2024/8/1
N2 - Aseptic loosening is the primary cause of failure following posterior-stabilized total knee arthroplasty. It is unclear whether tibial post loading of posterior-stabilized prosthesis increases the risk of aseptic loosening of the tibial prosthesis. The purpose of this study is to investigate the biomechanical effects of tibial post loading on the tibial prosthesis fixation interface during level walking, squatting, stair descent, and standing up-sitting down activities. In this paper, finite element models with and without post were established to compare the effects of tibial post loading on the von Mises stress of the proximal tibia, shear stress of the cement, and the bone-prosthesis interface micromotion during four physiological activities. The tibial post loading had an insignificant influence on tibial biomechanics and bone-prosthesis interface micromotion during leveling walking activity. However, compared to the insert without post condition, tibial post loading significantly increased the maximum tibial von Mises stress, the maximum shear stress in the medial of cement, and the bone-prosthesis interface peak micromotion by 912.84%, 612.77%, and 921.09%, respectively, at the moment of the maximum flexion angle for the stair descent activity, and 637.92%, 351.43%, and 519.13%, respectively, at the moment of the maximum flexion angle for the standing up-sitting down activity. Tibial post loading increased the risk of postoperative aseptic loosening of tibial prosthesis in patients with posterior-stabilized total knee arthroplasty, and it was recommended that the post-cam contact mechanism of posterior-stabilized prosthesis should be optimized to reduce the biomechanical impact of tibial post loading on tibial prosthesis fixation.
AB - Aseptic loosening is the primary cause of failure following posterior-stabilized total knee arthroplasty. It is unclear whether tibial post loading of posterior-stabilized prosthesis increases the risk of aseptic loosening of the tibial prosthesis. The purpose of this study is to investigate the biomechanical effects of tibial post loading on the tibial prosthesis fixation interface during level walking, squatting, stair descent, and standing up-sitting down activities. In this paper, finite element models with and without post were established to compare the effects of tibial post loading on the von Mises stress of the proximal tibia, shear stress of the cement, and the bone-prosthesis interface micromotion during four physiological activities. The tibial post loading had an insignificant influence on tibial biomechanics and bone-prosthesis interface micromotion during leveling walking activity. However, compared to the insert without post condition, tibial post loading significantly increased the maximum tibial von Mises stress, the maximum shear stress in the medial of cement, and the bone-prosthesis interface peak micromotion by 912.84%, 612.77%, and 921.09%, respectively, at the moment of the maximum flexion angle for the stair descent activity, and 637.92%, 351.43%, and 519.13%, respectively, at the moment of the maximum flexion angle for the standing up-sitting down activity. Tibial post loading increased the risk of postoperative aseptic loosening of tibial prosthesis in patients with posterior-stabilized total knee arthroplasty, and it was recommended that the post-cam contact mechanism of posterior-stabilized prosthesis should be optimized to reduce the biomechanical impact of tibial post loading on tibial prosthesis fixation.
KW - Total knee arthroplasty
KW - bone-prosthesis interface micromotion
KW - finite element analysis
KW - posterior-stabilized prosthesis
KW - stress distribution
KW - tibial post loading
UR - https://www.scopus.com/pages/publications/85201930648
U2 - 10.1177/09544119241272756
DO - 10.1177/09544119241272756
M3 - 文章
C2 - 39177049
AN - SCOPUS:85201930648
SN - 0954-4119
VL - 238
SP - 886
EP - 896
JO - Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
JF - Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
IS - 8-9
ER -