The simulation calibration technology of disk source for standard source efficiency of HPGe detector

  • Tsinghua University
  • Northwest Institute of Nuclear Technology

科研成果: 期刊稿件文章同行评审

2 引用 (Scopus)

摘要

Standard surface source was used to simulate the layer of different height in the standard source. The detection efficiency of (75 mm×25 mm241Am standard source and its gaseous source can be obtained by integration, the simulation calibration results are consistent with the calculation efficiency value of LabSOCS in 7%. The disk source experiments are also completed by LabSOCS efficiency calculation software, The detection efficiency of (75 mm×25 mm133Xe gaseous source and 137Cs liquid source, the simulation calibration results are consistent with the calculated efficiency value by LabSOCS within 1%. Therefore, the simulation calibration technology of disk source is identified to be correct and effective. Comparing with the calibration technology of standard source, less sources are required and the time spent can be much reduced, which will improve the work efficiency.

源语言英语
页(从-至)130-134
页数5
期刊Hedianzixue Yu Tance Jishu/Nuclear Electronics and Detection Technology
33
2
出版状态已出版 - 2月 2013
已对外发布