摘要
Microcalcification is one of the earliest and main indicators of breast cancer. Because dual-energy digital mammography could suppress the contrast between the adipose and glandular tissues of the breast, it is considered a promising technique that will improve the detection of microcalcification. In dual-energy digital mammography, the imaged object is a human breast, while in calibration measurements only the phantoms of breast tissue equivalent materials are available. Consequently, the differences between phantoms and breast tissues will lead to calibration phantom errors. Based on the dual-energy imaging model, formulae of calibration phantom errors are derived in this paper. Then, this type of error is quantitatively analyzed using publicly available data and compared with other types of error. The results demonstrate that the calibration phantom error is large and dominant in dual-energy mammography, seriously decreasing calculation precision. Further investigations on the physical meaning of calibration phantom error reveal that the imaged objects with the same glandular ratio have identical calibration phantom error. Finally, an error correction method is proposed based on our findings.
| 源语言 | 英语 |
|---|---|
| 页(从-至) | 6321-6336 |
| 页数 | 16 |
| 期刊 | Physics in Medicine and Biology |
| 卷 | 53 |
| 期 | 22 |
| DOI | |
| 出版状态 | 已出版 - 20 11月 2008 |
联合国可持续发展目标
此成果有助于实现下列可持续发展目标:
-
可持续发展目标 3 良好健康与福祉
学术指纹
探究 'The impact of calibration phantom errors on dual-energy digital mammography' 的科研主题。它们共同构成独一无二的指纹。引用此
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver