TY - JOUR
T1 - The Feasibility Study of Intraoperative RLN Monitoring Using Cricothyroid Membrane-Inserted Needle Electrodes During Thyroid Surgery
AU - Xu, Chongwen
AU - Wang, Xiaoxia
AU - Liu, Junsong
AU - Yan, Jinfeng
AU - Zhang, Shaoqiang
AU - Zheng, Zhiwei
AU - Zhao, Qian
AU - Feng, Yi
AU - Bai, Yanxia
AU - Zhao, Ruimin
AU - Li, Honghui
AU - Yao, Xiaobao
N1 - Publisher Copyright:
© 2023 American Academy of Otolaryngology–Head and Neck Surgery Foundation.
PY - 2023/9
Y1 - 2023/9
N2 - Objective: This study evaluated the feasibility, stability, safety, and economy of cricothyroid membrane (CM)-inserted needle electrodes for recurrent laryngeal nerve monitoring. Study Design: Parallel and controlled study. Setting: Clinical research center for thyroid diseases of Shaanxi province. Methods: A total of 64 patients in the needle electrodes group (104 recurrent laryngeal nerves [RLNs]) and 44 patients in the endotracheal tube (ETT)-based electrodes group (80 RLNs) underwent monitored thyroidectomy. The evoked electromyography (EMG) signals detected by the 2 electrodes were recorded and analyzed. The changes in EMG during Berry's ligament traction and tracheal displacement were compared. All patients underwent preoperative and postoperative laryngoscopy within 1 week. Results: Both electrodes successfully recorded typical evoked laryngeal EMG waveforms from RLNs. The needle electrodes recorded relatively higher amplitudes and similar latencies compared to ETT-based electrodes. The evoked EMG signals attributed to needle electrodes could accurately predict the function of RLNs with 100% sensitivity and specificity. The reduction in the recorded amplitudes attributed to needle electrodes was higher than that observed with ETT-based electrodes during Berry's ligament traction or trachea displacement, whereas a similar increase in the latencies was recorded in the 2 groups. Particularly, Berry's ligament traction was more likely to lead to EMG amplitude reduction and latency prolongation. The needle electrodes group recorded 2 cases of minor bleeding on the CM. The needle electrodes were more cost-effective than ETT-based electrodes. Conclusion: The CM-inserted needle electrodes are feasible, stable, safe, and economical for RLN monitoring, and they provide an alternative novel intraoperative neural monitoring format for thyroid surgeons.
AB - Objective: This study evaluated the feasibility, stability, safety, and economy of cricothyroid membrane (CM)-inserted needle electrodes for recurrent laryngeal nerve monitoring. Study Design: Parallel and controlled study. Setting: Clinical research center for thyroid diseases of Shaanxi province. Methods: A total of 64 patients in the needle electrodes group (104 recurrent laryngeal nerves [RLNs]) and 44 patients in the endotracheal tube (ETT)-based electrodes group (80 RLNs) underwent monitored thyroidectomy. The evoked electromyography (EMG) signals detected by the 2 electrodes were recorded and analyzed. The changes in EMG during Berry's ligament traction and tracheal displacement were compared. All patients underwent preoperative and postoperative laryngoscopy within 1 week. Results: Both electrodes successfully recorded typical evoked laryngeal EMG waveforms from RLNs. The needle electrodes recorded relatively higher amplitudes and similar latencies compared to ETT-based electrodes. The evoked EMG signals attributed to needle electrodes could accurately predict the function of RLNs with 100% sensitivity and specificity. The reduction in the recorded amplitudes attributed to needle electrodes was higher than that observed with ETT-based electrodes during Berry's ligament traction or trachea displacement, whereas a similar increase in the latencies was recorded in the 2 groups. Particularly, Berry's ligament traction was more likely to lead to EMG amplitude reduction and latency prolongation. The needle electrodes group recorded 2 cases of minor bleeding on the CM. The needle electrodes were more cost-effective than ETT-based electrodes. Conclusion: The CM-inserted needle electrodes are feasible, stable, safe, and economical for RLN monitoring, and they provide an alternative novel intraoperative neural monitoring format for thyroid surgeons.
KW - cricothyroid membrane
KW - intraoperative neuromonitoring
KW - needle electrodes
KW - recurrent laryngeal nerve
UR - https://www.scopus.com/pages/publications/85152800103
U2 - 10.1002/ohn.338
DO - 10.1002/ohn.338
M3 - 文章
C2 - 37051891
AN - SCOPUS:85152800103
SN - 0194-5998
VL - 169
SP - 589
EP - 597
JO - Otolaryngology - Head and Neck Surgery (United States)
JF - Otolaryngology - Head and Neck Surgery (United States)
IS - 3
ER -