跳到主要导航 跳到搜索 跳到主要内容

The different cones combination enhanced sensitivity on MC-ICP-MS: The results from boron isotope analysis

  • CAS - Institute of Earth Environment
  • Shaanxi Provincial Key Laboratory of Accelerator Mass Spectrometry and Application
  • National Institute of Metrology China
  • CAS - Institute of Geochemistry

科研成果: 期刊稿件文章同行评审

30 引用 (Scopus)

摘要

The sensitivity of multi-collector inductively coupled plasma mass spectrometers (MC-ICP-MS) can be significantly improved using different sample and skimmer cones with different orifice diameters and angles. In this study, an MC-ICP-MS with three typical combinations of sample and skimmer cones [(H skimmer cone + standard sample cone (Standard + H), X skimmer cone + standard sample cone (Standard + X) and X skimmer cone + Jet sample cone (Jet + X)] was used to measure the boron isotopes, and the effect of the different combinations of sample and skimmer cones on the sensitivity of the MC-ICP-MS and the underlying physical principles were investigated. The results suggest that more ions can be introduced into sample cones with larger orifices, which results in increased sensitivity. Compared with standard skimmer cones (which have cylindrical entrances and trumpet-shaped exits), X skimmer cones (which have trumpet-shaped entrances and exits) can provide more space for ion beams to entrance the skimmer's orifice after they are subjected to the space-charge effect.

源语言英语
页(从-至)33-37
页数5
期刊International Journal of Mass Spectrometry
408
DOI
出版状态已出版 - 10 9月 2016
已对外发布

学术指纹

探究 'The different cones combination enhanced sensitivity on MC-ICP-MS: The results from boron isotope analysis' 的科研主题。它们共同构成独一无二的学术指纹。

引用此