跳到主要导航 跳到搜索 跳到主要内容

The 3MV multi-element AMS in Xi'an, China: Unique features and preliminary tests

  • CAS - Institute of Earth Environment
  • Xi'an Jiaotong University

科研成果: 期刊稿件文章同行评审

71 引用 (Scopus)

摘要

A 3MV multi-element accelerator mass spectrometer (AMS) has been installed in Xi'an, China, and preliminary tests have been completed. The results of both background and precision tests for 4 nuclides are 3.1 × 10-16, 0.2% (14C); 1.8 × 10-14, 1.4% (10Be); 2.3 × 10-15, 1.14% (26Al); and 2.0 × 10-14, 1.75% (129I). The unique features of this facility are the newly developed ion source accepting solid and CO2 samples; the specially designed low-energy injector, including a "beam blanking unit" and "Q-snout"; the acceleration tube structure with the combined magnetic and electrostatic suppression; and the function of the slit stabilization in the post-acceleration system. These features are discussed in terms of the end-user's point of view.

源语言英语
页(从-至)285-293
页数9
期刊Radiocarbon
48
2
DOI
出版状态已出版 - 2006

学术指纹

探究 'The 3MV multi-element AMS in Xi'an, China: Unique features and preliminary tests' 的科研主题。它们共同构成独一无二的指纹。

引用此