Studies on injector pump-ultrasonic nebulizer feeding sample system for MPT-AES

  • Jilin University

科研成果: 期刊稿件文章同行评审

1 引用 (Scopus)

摘要

The feeding-sample system used by microwave plasma torch atomic emission spectroscopy (MPT-AES) is the pneumatic nebulization system; its efficiency, however, is not good. A newly built injector pump-ultrasonic nebulizer combination feeding-sample system has been designed. Its performance was tested and compared with that of the pneumatic nebulization system. It can be concluded that the newly built feeding-sample system can increase the spectral line intensity by about two to three times and decrease the detection limit by about 2 to 10 times. Moreover, this newly built system can reduce the time taken washing the sample cell from 30 min or so to about 10 min.