跳到主要导航 跳到搜索 跳到主要内容

Silicon carbide Schottky diode detector for fusion neutron yields monitoring using associated particle method

  • Pengying Wan
  • , Linyue Liu
  • , Runlong Gao
  • , Yangbo Nie
  • , Liang Chen
  • , Xiao Ouyang
  • , Xianpeng Zhang
  • , Hui Li
  • , Kai Zhang
  • , Xichao Ruan
  • , Xiaoping Ouyang
  • Xi'an Jiaotong University
  • Northwest Institute of Nuclear Technology
  • Sun Yat-Sen University
  • China National Nuclear Corporation
  • Beijing Normal University

科研成果: 期刊稿件 › 文章 › 同行评审

5 引用 (Scopus)

摘要

The associated particle method is proven to be one of ideal options for D-T fusion neutron detection, acquiring accurate neutron yields by monitoring α particle. Semiconductor detectors can be used to detect α particle for compact size, high energy-resolution, and stable response. SiC detector with high radiation resistance is benefit for long service time in fusion neutron monitoring. In this paper, a SiC Schottky diode detector is used to detect associated α particles for D-T fusion neutron monitoring, and successfully gets accurate neutron yields, being highly consistent with the conventional Si surface barrier detector. It is encouraging to find out SiC detector can afford 2.56 × 109 cm−2 α particles irradiation, with slight performance degradation of dark current and energy resolution, indicating that SiC detector is a promising candidate for fusion neutron detection.