摘要
A reverse flow-injection spectrophotometric method with diphenyl carbazide as developer and sulfuric acid as medium for determination of Cr(VI) in electroplating waste water and sludge has been developed. Experiment conditions of reverse flow-injection analysis are optimized. The optimal conditions were as follows: the concentration of diphenylcarbazide in color solution was 0.5 mmol/L, the acidity of sulfuric acid of reaction medium was 0.5 mol/L, the length of reaction coil was 200 cm, the flow of carrying solution (blank solution) was 0.8 mL/min, sample flow was 1.6 mL/min and the injection dosage of developer reagent was 100 μL. Under the optimal injection conditions, the detection limit of this method was 0.5 μg/L with the linear range of 1-100 μg/L. The results of this method used for the determination of Cr(VI) in standard water sample were in accordant with the certified values. This method also could be used to determine Cr(VI) in electroplating waste water and sludge, and the results were in agreement with those obtained by AAS.
| 源语言 | 英语 |
|---|---|
| 页(从-至) | 47-50 |
| 页数 | 4 |
| 期刊 | Yejin Fenxi/Metallurgical Analysis |
| 卷 | 28 |
| 期 | 4 |
| 出版状态 | 已出版 - 4月 2008 |
| 已对外发布 | 是 |
学术指纹
探究 'Reverse flow-inject ion spectrophotometric determination of chromium (VI) in electroplating waste water and sludge' 的科研主题。它们共同构成独一无二的学术指纹。引用此
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver