摘要
The study demonstrates the applicability of laser ionization time-of-flight mass spectrometryfor real-time measurement of polychlorinated biphenyls (PCBs). Picosecond 266-nm laser light ionization reduced fragmentation and provided very high PCB detection sensitivity. This high sensitivity has advantages in terms of real-time monitoring capability as compared to the conventional GC-ECD or GC-MS methods, which require at least several days for the analysis of PCBs. Detection sensitivity of under 0.01 mg/Nm3 was achieved with a 1-min measuring time; this sensitivity is superior to the exhaust gas control guideline of 0.15 mg/Nm3 by a factor of 10. A prototype PCB monitoring device has been developed and tested in a pilot PCB treatment plant. The 1-min detection time represents a substantial advance in the monitoring of exhaust gas and the workplace atmosphere in accordance with safety regulations.
| 源语言 | 英语 |
|---|---|
| 页(从-至) | 4737-4742 |
| 页数 | 6 |
| 期刊 | Environmental Science and Technology |
| 卷 | 37 |
| 期 | 20 |
| DOI | |
| 出版状态 | 已出版 - 15 10月 2003 |
| 已对外发布 | 是 |
学术指纹
探究 'Real-time PCB monitoring using time-of-flight mass spectrometry with picosecond laser ionization' 的科研主题。它们共同构成独一无二的指纹。引用此
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver