跳到主要导航 跳到搜索 跳到主要内容

Principle and application of sub-threshold correcting for sensitivity calibration by single particle

  • Xin Jian Tan
  • , Xiao Ping Ouyang
  • , Zhao Hui Song
  • , Xi Chao Ruan
  • , Xiao Chuan Zhao
  • , Qun Shu Wang
  • , Ke Jun Kang
  • Tsinghua University
  • Northwest Institute of Nuclear Technology
  • China National Nuclear Corporation

科研成果: 期刊稿件文章同行评审

摘要

The technique of calibration by single particle is very important for the current detector on the energy response research. Based on the analysis of the energy response characteristic and pulsed signal producing process, a method was proposed for the sub-threshold correcting, which extended the calibrating technique to much lower energy range. According to the correcting technique, the measured sensitivities to 60Co by single particle calibration and traditional current calibration match each other within 1.1% difference. The sensitivity ratio of 60Co to 137Cs matches the theoretical value within 7.0% difference.

源语言英语
页(从-至)351-355
页数5
期刊Yuanzineng Kexue Jishu/Atomic Energy Science and Technology
45
3
出版状态已出版 - 3月 2011
已对外发布

学术指纹

探究 'Principle and application of sub-threshold correcting for sensitivity calibration by single particle' 的科研主题。它们共同构成独一无二的指纹。

引用此