This paper presents a portable miniaturized integration UV-VIS absorbance detection system for the microchip electrophoresis (MCE). The TOCON_nano high sensitive SiC pre-amplified UV Sensor (Sglux GmbH, Ostendstrasse, Berlin, Germany) is used to measure extremely low UV radiation intensities in the nanowatt scale. This detection system volume is 280 ×200×150 mm3, the weight is less than 2Kg. The PMMA was used as the MCE substrate materials. And the electrophoresis microchannels are fabricated with the CNC milling, have the cross section area of 150μm ×150μm. With this detection system, the detection limit for the Cl- electrophoresis can reach 10-10mol at the indirect UV-VIS detection mode. Two inorganic ions Cl- and SO4 2- could reach the baseline separation in 3 min.