Optimal design for high-field MRI superconducting magnet

  • Chunzhong Wang
  • , Tianqing Chang
  • , Ming Rong
  • , Yinming Dai
  • , Zhipeng Ni
  • , Lankai Li
  • , Qiuliang Wang

科研成果: 期刊稿件文章同行评审

16 引用 (Scopus)

摘要

An optimal design method for high-field MRI magnet is put forward. The design procedure can be divided into two steps. First step is to design a set of coils to achieve the necessary center field, which is called the background field, and it is based on the optimal algorithm of minimum volume consumption by SQP. During the optimization, the safety factor of superconducting wires has been considered. The second step is to design a set of correction coils to achieve the necessary homogeneity, and the magnetic field produced by previous coilsis set as background field, then optimize the homogeneity based on the nonlinear Least Square Method.

源语言英语
文章编号5659982
页(从-至)2245-2249
页数5
期刊IEEE Transactions on Applied Superconductivity
21
3 PART 2
DOI
出版状态已出版 - 6月 2011

学术指纹

探究 'Optimal design for high-field MRI superconducting magnet' 的科研主题。它们共同构成独一无二的指纹。

引用此