跳到主要导航 跳到搜索 跳到主要内容

Nanocrown electrodes for parallel and robust intracellular recording of cardiomyocytes

  • Zeinab Jahed
  • , Yang Yang
  • , Ching Ting Tsai
  • , Ethan P. Foster
  • , Allister F. McGuire
  • , Huaxiao Yang
  • , Aofei Liu
  • , Csaba Forro
  • , Zen Yan
  • , Xin Jiang
  • , Ming Tao Zhao
  • , Wei Zhang
  • , Xiao Li
  • , Thomas Li
  • , Annalisa Pawlosky
  • , Joseph C. Wu
  • , Bianxiao Cui
  • Stanford University
  • University of California at San Diego
  • University of North Texas
  • LLC
  • Nationwide Children’s Hospital
  • Ohio State University
  • Alphabet Inc.

科研成果: 期刊稿件 › 文章 › 同行评审

65 引用 (Scopus)

摘要

Drug-induced cardiotoxicity arises primarily when a compound alters the electrophysiological properties of cardiomyocytes. Features of intracellular action potentials (iAPs) are powerful biomarkers that predict proarrhythmic risks. In the last decade, a number of vertical nanoelectrodes have been demonstrated to achieve parallel and minimally-invasive iAP recordings. However, the large variability in success rate and signal strength have hindered nanoelectrodes from being broadly adopted for proarrhythmia drug assessment. In this work, we develop vertically-aligned nanocrown electrodes that are mechanically robust and achieve > 99% success rates in obtaining intracellular access through electroporation. We validate the accuracy of nanocrown electrode recordings by simultaneous patch clamp recording from the same cell. Finally, we demonstrate that nanocrown electrodes enable prolonged iAP recording for continual monitoring of the same cells upon the sequential addition of four incremental drug doses. Our technology development provides an advancement towards establishing an iAP screening assay for preclinical evaluation of drug-induced arrhythmogenicity.

源语言英语
期刊论文编号2253
期刊Nature Communications
卷13
期1
DOI
出版状态已出版 - 12月 2022
已对外发布是

学术指纹

探究 'Nanocrown electrodes for parallel and robust intracellular recording of cardiomyocytes' 的科研主题。它们共同构成独一无二的学术指纹。

引用此