跳到主要导航 跳到搜索 跳到主要内容

Multiple test zones for improved detection performance in lateral flow assays

  • Xi'an Jiaotong University
  • University of Malaya
  • Zhejiang University
  • Collaborative Innovation Centre for Diagnosis and Treatment of Infectious Diseases

科研成果: 期刊稿件文章同行评审

68 引用 (Scopus)

摘要

Lateral flow assays (LFAs) have found applications in clinical diagnostics, veterinary medicine, food safety, environmental monitoring, bio-defence, drug abuse, etc. However, conventional LFAs with single test zone for one analyte (sLFA) show qualitative, positive/negative results (that are usually not so accurate for decision-making) and suffer from the hook effect (that easily induces false negative results). In this study, a modified LFA with multiple test zones for one analyte (mLFA) is developed for detection of nucleic acid showing two methods to interpret assay results. One is the sum of detection signals of test zones, which exhibits a stable and broader detection range compared with sLFA. The other is the pattern of test zones that is presented is correlated to the concentration of analyte, facilitating semi-quantitative detection without the aid of a reader or analyzer.

源语言英语
页(从-至)484-488
页数5
期刊Sensors and Actuators B: Chemical
243
DOI
出版状态已出版 - 2017

联合国可持续发展目标

此成果有助于实现下列可持续发展目标:

  1. 可持续发展目标 3 - 良好健康与福祉
    可持续发展目标 3 良好健康与福祉

学术指纹

探究 'Multiple test zones for improved detection performance in lateral flow assays' 的科研主题。它们共同构成独一无二的学术指纹。

引用此