MDM-PCCT: Multiple Dynamic Modulations for High-Performance Spectral PCCT Imaging

  • Danyang Li
  • , Dong Zeng
  • , Sui Li
  • , Yongshuai Ge
  • , Zhaoying Bian
  • , Jing Huang
  • , Jianhua Ma

科研成果: 期刊稿件文章同行评审

11 引用 (Scopus)

摘要

Photon counting computed tomography (PCCT) has the ability to identify individual photons, resulting in quantitative material identification. Meanwhile, several technical challenges still exist in current PCCT imaging systems, including increased noise and suboptimal bin selection. These nonideal effects can substantially degrade the reconstruction performance and material estimation accuracy. To address these issues, in this work, we present a novel system for high-performance spectral PCCT imaging, which is a combination of multiple dynamic modulations, interpolation-based measurements processing strategy and advanced reconstruction method. For simplicity, this new PCCT imaging system is referred to as "MDM-PCCT". Specifically, the multiple dynamic modulations consist of dynamic kVp modulation, dynamic spectrum modulation and dynamic energy threshold modulation. In the dynamic kVp modulation, three kVp values, i.e., 80, 110 and 140, are included, and the tube voltage waveform follows a sinusoidal curve which is more practical than the rectangular curve in the fast kV switching mode. In the dynamic spectrum modulation, the X-ray spectra are processed by selective spatial-spectral filters to balance the X-ray fluxes and increase the spectral separation. In the dynamic energy threshold modulation, the energy threshold is adaptively changed to determine the optimal bin selection. Furthermore, we propose an energy threshold determination method and interpolation-based measurements processing strategy to address the issue of non-uniform and sparse-view PCCT measurements, respectively. In addition, by considering the intrinsic characteristics of the MDM-PCCT images, we utilize an enhanced total variation regularized model for images reconstruction. Finally, numerical and preclinical studies demonstrate that the presented MDM-PCCT imaging system is capable of yielding uniform and high-fidelity PCCT measurements with noise consistency, and the presented reconstruction method further improves the image quality and material decomposition accuracy.

源语言英语
页(从-至)3630-3642
页数13
期刊IEEE Transactions on Medical Imaging
39
11
DOI
出版状态已出版 - 11月 2020
已对外发布

学术指纹

探究 'MDM-PCCT: Multiple Dynamic Modulations for High-Performance Spectral PCCT Imaging' 的科研主题。它们共同构成独一无二的指纹。

引用此