跳到主要导航 跳到搜索 跳到主要内容

Low temperature plasma ablation device for minimally invasive surgery

  • Xi'an Jiaotong University
  • Chang'an University

科研成果: 期刊稿件文章同行评审

7 引用 (Scopus)

摘要

This study introduces the design principle of a high-frequency square wave power supply, as well as the simulation and design process of a minimally invasive electrosurgical scalpel, which formed the low temperature plasma in saline solution. In order to verify the performance of the device, saline discharge experiments and pork ablation experiments were performed. An alternating high-frequency square-wave voltage was applied between coaxial double-layer electrodes with a spacing of 400 μm. The waveforms and amplitude of output voltage and current, conductance between electrodes, and output power were measured in saline discharge experiments. The results of pork ablation experiments with different power sources demonstrate that the device can generate low-temperature plasma for ablation rather than relying on thermal effects. The device can ablate pork under the driving of a 10 V, 100 kHz square wave voltage. In addition, the ablation speed and area will be greater with an increase in voltage or frequency.

源语言英语
期刊论文编号014101
期刊Review of Scientific Instruments
94
1
DOI
出版状态已出版 - 1 1月 2023

学术指纹

探究 'Low temperature plasma ablation device for minimally invasive surgery' 的科研主题。它们共同构成独一无二的学术指纹。