Iterative scatter correction for x-ray cone-beam CT with semi-transparent beam stop array

  • Xi'an Jiaotong University

科研成果: 书/报告/会议事项章节会议稿件同行评审

6 引用 (Scopus)

摘要

X-ray scatter correction is an open problem in the research of cone beam CT (CBCT). In this paper, using semitransparent beam stop array (BSA) introduced recently1, an improved BSA method2 is proposed. A simple iterative correction algorithm is developed based on the knowledge that mostly primary varies continuously. With the proposed method, scatter correction could be accomplished without extra scan. Experiments are carried out based on the Monte Carlo (MC) simulation (EGSnrc), the mean square error (MSE) is reduced from 50.71% (with no correction) to 2.54%, and there is no visible negative impact introduced by the correction.

源语言英语
主期刊名Medical Imaging 2009
主期刊副标题Physics of Medical Imaging
DOI
出版状态已出版 - 2009
活动Medical Imaging 2009: Physics of Medical Imaging - Lake Buena Vista, FL, 美国
期限: 9 2月 200912 2月 2009

丛书

姓名Progress in Biomedical Optics and Imaging - Proceedings of SPIE
7258
ISSN(印刷版)1605-7422

会议

会议Medical Imaging 2009: Physics of Medical Imaging
国家/地区美国
Lake Buena Vista, FL
时期9/02/0912/02/09