TY - JOUR
T1 - Investigation into factors affecting the mechanical behaviours of a patient-specific vertebral body replacement
AU - Wang, Ling
AU - Kang, Jianfeng
AU - Shi, Lei
AU - Fu, Jun
AU - Li, Dichen
AU - Guo, Zheng
AU - Liu, Chaozong
AU - Dong, Shuangpeng
AU - Jiang, Xi
N1 - Publisher Copyright:
© 2018, © IMechE 2018.
PY - 2018/4/1
Y1 - 2018/4/1
N2 - Most vertebral body implants that are currently designed and produced in batches have difficulty meeting the patient-specific demands. Moreover, several complications, including a low fusion rate, subsidence occurrence, and rod displacement, are associated with these implants. This study aims to investigate the effects of patient-specific geometric and clinical parameters on the biomechanics of a vertebral body replacement. A three-dimensional patient-specific vertebral body replacement model was established as the basic model for parametric studies, including the anatomic design of the endplates, tilting angle, thickness, and dislocation of the vertebral body implant. A finite element analysis was applied to determine the stress distribution of the vertebral body implant when under various loading conditions. The model with an anatomical interfacing design generates 75% less stress concentration compared to a flat design; the peak stress of the model with a tilted angle closely matching the replaced vertebra segment is decreased by 30%; and the thickness close to the cortical bone can offer better bone growth capability and long-term stability. Patient-specific geometrical parameters were found to significantly affect the biomechanics of a vertebral body replacement, and therefore, a design customized especially for the endplates is necessary for better stability and long-term longevity of the prostheses. Regardless of such progress, how to balance the stability of a vertebral body implant and the safety of the peripheral nervous system remains a clinical challenge.
AB - Most vertebral body implants that are currently designed and produced in batches have difficulty meeting the patient-specific demands. Moreover, several complications, including a low fusion rate, subsidence occurrence, and rod displacement, are associated with these implants. This study aims to investigate the effects of patient-specific geometric and clinical parameters on the biomechanics of a vertebral body replacement. A three-dimensional patient-specific vertebral body replacement model was established as the basic model for parametric studies, including the anatomic design of the endplates, tilting angle, thickness, and dislocation of the vertebral body implant. A finite element analysis was applied to determine the stress distribution of the vertebral body implant when under various loading conditions. The model with an anatomical interfacing design generates 75% less stress concentration compared to a flat design; the peak stress of the model with a tilted angle closely matching the replaced vertebra segment is decreased by 30%; and the thickness close to the cortical bone can offer better bone growth capability and long-term stability. Patient-specific geometrical parameters were found to significantly affect the biomechanics of a vertebral body replacement, and therefore, a design customized especially for the endplates is necessary for better stability and long-term longevity of the prostheses. Regardless of such progress, how to balance the stability of a vertebral body implant and the safety of the peripheral nervous system remains a clinical challenge.
KW - Vertebral body replacement
KW - bone graft fusion
KW - finite element analysis
KW - patient-specific design
KW - subsidence
UR - https://www.scopus.com/pages/publications/85045297030
U2 - 10.1177/0954411918754926
DO - 10.1177/0954411918754926
M3 - 文章
C2 - 29383959
AN - SCOPUS:85045297030
SN - 0954-4119
VL - 232
SP - 378
EP - 387
JO - Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
JF - Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
IS - 4
ER -