TY - JOUR
T1 - Interdevice comparison of retinal sensitivity assessments in a healthy population
T2 - The CenterVue MAIA and the Nidek MP-3 microperimeters
AU - Balasubramanian, Siva
AU - Uji, Akihito
AU - Lei, Jianqin
AU - Velaga, Swetha
AU - Nittala, Muneeswar
AU - Sadda, Srini Vas
N1 - Publisher Copyright:
© Article author(s) (or their employer(s) unless otherwise stated in the text of the article) 2018. All rights reserved.
PY - 2018/1/1
Y1 - 2018/1/1
N2 - Background To compare and correlate the retinal sensitivity measurements obtained with Nidek Microperimetry-3 (MP-3) and the CenterVue Macular Integrity Assessment (MAIA) microperimeters among healthy subjects. Methods In this prospective comparative study, 31 eyes of 23 subjects underwent complete ophthalmological examination including retinal sensitivity assessments using two microperimeters, the MP-3 (Nidek Technologies) and the MAIA (CenterVue). The mean retinal sensitivity (dB) and its corresponding luminance (asb) and contrast (log units) were analysed between the two instruments. The interdevice reproducibility and level of agreement between the sensitivity values of the devices were assessed. Results The mean retinal sensitivity (dB) measured by the MP-3 (25.02±1.06 dB, range: 20.90-26.70) was significantly (p 0.0001) lower compared with the MAIA (30.68±0.74 dB, range: 28-31.84). The luminosity levels were significantly (p 0.0001) higher with the MP3 (7.75±1.31 asb, range: 6.44-9.06) compared with the MAIA (0.92±0.14 asb, range: 0.78-1.06). The contrast sensitivity was significantly higher for the MP-3 (0.94±0.33 log units, range: 0.61-1.27) compared with the MAIA (0.23±0.03 log units, range: 0.20-0.26). Despite these absolute differences, the intraclass coefficient was 0.85 (95% CI 0.70 to 0.92) between the two devices after applying a standard correction factor to each data point (MAIA sensitivity=MP-3 sensitivity+5.65) with a mean difference between MAIA and MP-3 of 0.01. Conclusion Retinal sensitivity measures higher, but luminance and contrast sensitivity measure lower for MAIA-generated values compared with the MP-3. The relationships, however, appeared fairly consistent, and application of a standard correction factor allowed the data to be inter-related, at least for normal eyes.
AB - Background To compare and correlate the retinal sensitivity measurements obtained with Nidek Microperimetry-3 (MP-3) and the CenterVue Macular Integrity Assessment (MAIA) microperimeters among healthy subjects. Methods In this prospective comparative study, 31 eyes of 23 subjects underwent complete ophthalmological examination including retinal sensitivity assessments using two microperimeters, the MP-3 (Nidek Technologies) and the MAIA (CenterVue). The mean retinal sensitivity (dB) and its corresponding luminance (asb) and contrast (log units) were analysed between the two instruments. The interdevice reproducibility and level of agreement between the sensitivity values of the devices were assessed. Results The mean retinal sensitivity (dB) measured by the MP-3 (25.02±1.06 dB, range: 20.90-26.70) was significantly (p 0.0001) lower compared with the MAIA (30.68±0.74 dB, range: 28-31.84). The luminosity levels were significantly (p 0.0001) higher with the MP3 (7.75±1.31 asb, range: 6.44-9.06) compared with the MAIA (0.92±0.14 asb, range: 0.78-1.06). The contrast sensitivity was significantly higher for the MP-3 (0.94±0.33 log units, range: 0.61-1.27) compared with the MAIA (0.23±0.03 log units, range: 0.20-0.26). Despite these absolute differences, the intraclass coefficient was 0.85 (95% CI 0.70 to 0.92) between the two devices after applying a standard correction factor to each data point (MAIA sensitivity=MP-3 sensitivity+5.65) with a mean difference between MAIA and MP-3 of 0.01. Conclusion Retinal sensitivity measures higher, but luminance and contrast sensitivity measure lower for MAIA-generated values compared with the MP-3. The relationships, however, appeared fairly consistent, and application of a standard correction factor allowed the data to be inter-related, at least for normal eyes.
KW - imaging
KW - retina
UR - https://www.scopus.com/pages/publications/85059048081
U2 - 10.1136/bjophthalmol-2017-310258
DO - 10.1136/bjophthalmol-2017-310258
M3 - 文章
C2 - 28495907
AN - SCOPUS:85059048081
SN - 0007-1161
VL - 102
SP - 109
EP - 113
JO - British Journal of Ophthalmology
JF - British Journal of Ophthalmology
IS - 1
ER -