跳到主要导航 跳到搜索 跳到主要内容

Influence of Acetabular Cup Rim Design on the Contact Stress During Edge Loading in Ceramic-on-Ceramic Hip Prostheses

  • University of Leeds

科研成果: 期刊稿件 › 文章 › 同行评审

43 引用 (Scopus)

摘要

The purpose of the study was to investigate the contact stresses in 3 different acetabular cup rim designs (new, worn, chamfer) during edge loading, after microseparation of ceramic on ceramic hip prostheses. A 3-dimensional finite element analysis was conducted for a 28-mm diameter alumina ceramic bearing with a radial clearance of 40 μm using a normal load of 2500 N under edge loading. At a separation distance of 250 μm, the maximum tensile stress in the "new" design was of similar magnitude to the flexural strength of the alumina material that supports the localized breakdown (stripe wear) of the acetabular cup surface observed clinically. Introducing a 2.5-mm radius chamfer should reduce the maximum tensile stress in the region of 60%.

源语言英语
页(从-至)131-136
页数6
期刊Journal of Arthroplasty
卷26
期1
DOI
出版状态已出版 - 1月 2011
已对外发布是

学术指纹

探究 'Influence of Acetabular Cup Rim Design on the Contact Stress During Edge Loading in Ceramic-on-Ceramic Hip Prostheses' 的科研主题。它们共同构成独一无二的学术指纹。

引用此