跳到主要导航 跳到搜索 跳到主要内容

Implementing a two-stage, shim field-calibrated superconducting shimming method on a 7 T cryogen-free small animal MRI magnet

  • Jinhao Liu
  • , Yaohui Wang
  • , Miutian Wang
  • , Wenchen Wang
  • , Gang Yang
  • , Weimin Wang
  • , Qiuliang Wang
  • , Feng Liu
  • Xi'an Jiaotong University
  • University of Chinese Academy of Sciences
  • CAS - Institute of Electrical Engineering
  • Peking University
  • University of Queensland
  • Shenzhen Bay Laboratory

科研成果: 期刊稿件 › 文章 › 同行评审

5 引用 (Scopus)

摘要

Ultrahigh field systems (≥ 7 T) can increase the signal-to-noise ratio of magnetic resonance imaging (MRI), improving imaging performance compared to systems with lower fields. However, these enhancements heavily rely on a high B0 magnetic field homogeneity level, which can be achieved through superconducting shimming. This paper presents a novel two-stage superconducting shimming method designed to achieve precise shimming for a 7 T MRI superconducting magnet. In the initial stage, detailed measurements and fittings were conducted to determine the current polarity and the axial or circumferential positions of the shim fields. Subsequently, an optimization strategy was implemented to determine the optimal shim currents with a flexible target field. The second stage involves an iterative process to fine-tune the current of a specific shim coil, identified as having the most significant impact on field homogeneity. The overall fitness of 99.5% underscores the precision in determining the current polarity and position of the shim fields. Significantly, the calibrated shim system substantially improves the peak-to-peak and Root Mean Square Error (RMSE) field homogeneities from 107.42 ppm and 37.00 ppm to 11.12 ppm and 3.26 ppm, respectively, representing improvements of 89.65% and 91.19%. Furthermore, the simulation results of the fine-tuning stage demonstrate additional enhancements in peak-to-peak field homogeneity, to 9.9 ppm by reducing the current of the Z2 shim coil by 51.3 mA. Additionally, the shimmed magnetic field exhibited high time stability, with a maximum variation of only 27 µT observed within 48 h. Thus, the proposed two-stage superconducting shimming framework effectively addresses the challenge of imperfect B0 magnetic fields, enhancing peak-to-peak and RMSE field homogeneity. The stepwise optimized approach also mitigates deviations caused by shim-to-shim coupling, demonstrating its efficacy in achieving precise shimming in ultrahigh-field MRI systems.

源语言英语
期刊论文编号107787
期刊Journal of Magnetic Resonance
卷368
DOI
出版状态已出版 - 11月 2024

学术指纹

探究 'Implementing a two-stage, shim field-calibrated superconducting shimming method on a 7 T cryogen-free small animal MRI magnet' 的科研主题。它们共同构成独一无二的学术指纹。

引用此