TY - JOUR
T1 - Histotripsy Using Fundamental and Second Harmonic Superposition Combined with Hundred-Microsecond Ultrasound Pulses
AU - Li, Yujiao
AU - Wang, Rui
AU - Lu, Mingzhu
AU - Zhang, Linglu
AU - Liu, Yanshan
AU - Han, Dan
AU - Wang, Xuan
AU - Geng, Yizhe
AU - Wan, Mingxi
N1 - Publisher Copyright:
© 2018 World Federation for Ultrasound in Medicine & Biology
PY - 2018/10
Y1 - 2018/10
N2 - A novel histotripsy approach based on fundamental and second harmonic superposition and incorporating hundred-microsecond-long pulses and two-stage pulse protocol is proposed in this study to rapidly generate mechanically homogenized lesions. Two pulse stages were applied: stage 1, pulses with a pulse duration of 500–600 μs and pulse repetition frequency of 100 Hz, and stage 2, multiple periods, each composed of multiple pulses with the same pulse duration and pulse repetition frequency as those in stage 1, but with an off-time of 600 ms between periods. A custom-designed 1.1/2.2-MHz two-element confocal-annular array, with an f-number of 0.69, and lateral and axial full width at half-maximum pressure dimensions of approximately 1.0 and 6.0 mm, was used. The peak positive/negative pressures at the focus were +22/–7 MPa for 1.1 MHz and +56/–14 MPa with shock wave for 2.2 MHz. To investigate the feasibility of this approach, experiments were designed and performed in tissue-mimicking polyacrylamide gel phantoms with bovine serum albumin and in ex vivo porcine tissues. Cavitation and boiling activities were observed through high-speed photography, and the corresponding acoustic emissions were recorded through passive cavitation detection. Ex vivo experimental results revealed that complete tissue homogeneous regions with regular long tear shape and typical dimensions of 5.80 ± 0.19 mm in axial and 2.20 ± 0.26 mm in lateral were successfully generated in porcine kidney samples. The hematoxylin and eosin staining evidenced that the lesions were thoroughly homogenized and sharply demarcated from untreated regions. These results indicated that the histotripsy approach using fundamental and second harmonic superposition combined with hundred-microsecond pulses and two-stage pulse protocol can efficiently obtain a mechanical disruption of soft tissues with spatial precision, and this approach may have the potential to be developed as a useful tool for precise tumor treatment.
AB - A novel histotripsy approach based on fundamental and second harmonic superposition and incorporating hundred-microsecond-long pulses and two-stage pulse protocol is proposed in this study to rapidly generate mechanically homogenized lesions. Two pulse stages were applied: stage 1, pulses with a pulse duration of 500–600 μs and pulse repetition frequency of 100 Hz, and stage 2, multiple periods, each composed of multiple pulses with the same pulse duration and pulse repetition frequency as those in stage 1, but with an off-time of 600 ms between periods. A custom-designed 1.1/2.2-MHz two-element confocal-annular array, with an f-number of 0.69, and lateral and axial full width at half-maximum pressure dimensions of approximately 1.0 and 6.0 mm, was used. The peak positive/negative pressures at the focus were +22/–7 MPa for 1.1 MHz and +56/–14 MPa with shock wave for 2.2 MHz. To investigate the feasibility of this approach, experiments were designed and performed in tissue-mimicking polyacrylamide gel phantoms with bovine serum albumin and in ex vivo porcine tissues. Cavitation and boiling activities were observed through high-speed photography, and the corresponding acoustic emissions were recorded through passive cavitation detection. Ex vivo experimental results revealed that complete tissue homogeneous regions with regular long tear shape and typical dimensions of 5.80 ± 0.19 mm in axial and 2.20 ± 0.26 mm in lateral were successfully generated in porcine kidney samples. The hematoxylin and eosin staining evidenced that the lesions were thoroughly homogenized and sharply demarcated from untreated regions. These results indicated that the histotripsy approach using fundamental and second harmonic superposition combined with hundred-microsecond pulses and two-stage pulse protocol can efficiently obtain a mechanical disruption of soft tissues with spatial precision, and this approach may have the potential to be developed as a useful tool for precise tumor treatment.
KW - Boiling
KW - Energetic cavitation
KW - Fundamental and second harmonic superposition
KW - Histotripsy
KW - Hundred-microsecond pulses
UR - https://www.scopus.com/pages/publications/85050281706
U2 - 10.1016/j.ultrasmedbio.2018.05.024
DO - 10.1016/j.ultrasmedbio.2018.05.024
M3 - 文章
C2 - 30054023
AN - SCOPUS:85050281706
SN - 0301-5629
VL - 44
SP - 2089
EP - 2104
JO - Ultrasound in Medicine and Biology
JF - Ultrasound in Medicine and Biology
IS - 10
ER -