Highly Shielded Gradient Coil Design for a Superconducting Planar MRI System

  • Yaohui Wang
  • , Qiuliang Wang
  • , Hongyi Qu
  • , Yang Liu
  • , Feng Liu
  • Chinese Academy of Sciences
  • University of Queensland

科研成果: 期刊稿件文章同行评审

15 引用 (Scopus)

摘要

In a planar, superconducting magnetic resonance imaging (MRI) system, the gradient assembly is placed into the groove of the SC magnet. Conventional gradient coil design method considers the shielding of pole face only, but neglects the surrounding metal structures at the coil side, thus leading to large stray field leakage and resulting in serious eddy current artefact. A novel coil shielding method was proposed in this work by a full consideration of the stray fields on the pole face and also the coil ends. The gradient coil design exemplification of a 0.7T planar superconducting MRI system was presented. In the new design, the maximum stray field at the surface of the ambient structure was reduced more than six times for the transverse coils and around four times for the longitudinal coil. The highly shielded gradient coil also produced linear gradient fields (<5%) over the imaging volume.