跳到主要导航 跳到搜索 跳到主要内容

High sensitive fission neutron detector system with a lead slot collimator

  • Xiao Ping Ouyang
  • , Zhen Fu Li
  • , Qun Shu Wang
  • , Yu Kun Huo
  • , Yan Liang Ma
  • , Qian Mei Zhang
  • , Guo Guang Zhang
  • , Yu Ren Jin
  • Tsinghua University
  • Fudan University

科研成果: 期刊稿件文章同行评审

11 引用 (Scopus)

摘要

A fission neutron detector system with frontward-extended lead slot collimator has been developed and tested. By designing the slot collimator and developing large area PIN detectors as well as large area fissionable target with thin Be backing, this detector system can reach high neutron sensitivities up to 10-16C · cm2, which is about 4-5 orders of magnitude higher than that of existing fission detectors, and high n/γ discrimination ratio up to 30.

源语言英语
页(从-至)4643-4647
页数5
期刊Wuli Xuebao/Acta Physica Sinica
54
10
DOI
出版状态已出版 - 10月 2005