跳到主要导航 跳到搜索 跳到主要内容

Flow injection chemiluminescence determination of rutin based on its enhancing effect on the luminol-ferricyanide/ferrocyanide system

  • Shaanxi Normal University

科研成果: 期刊稿件 › 文章 › 同行评审

21 引用 (Scopus)

摘要

A new flow injection chemiluminescence (CL) method has been developed for determination of rutin. It is based on the direct CL from the reaction of rutin with luminol in the presence of potassium ferricyanide and potassium ferrocyanide in alkaline solution. The linear range for the determination of rutin is 0.1-2 μg/ml and the detection limit is 0.03 μg/ml rutin. The relative standard deviation is 0.8% for 0.4 μg/ml rutin solution in eleven repeated measurements. The method has been successfully applied to the determination of rutin in commercial rutin tablets. The reaction mechanism has also been discussed briefly.

源语言英语
页(从-至)1741-1748
页数8
期刊Analytical Letters
卷34
期10
DOI
出版状态已出版 - 2001
已对外发布是

学术指纹

探究 'Flow injection chemiluminescence determination of rutin based on its enhancing effect on the luminol-ferricyanide/ferrocyanide system' 的科研主题。它们共同构成独一无二的学术指纹。

引用此