Dynamic calibration method for detection efficiency of radioactive source

  • Zining Tian
  • , Xiaoping Ouyang
  • , Bin Han
  • , Wei Chen
  • , Chuanying Su
  • , Wenbiao Liu
  • , Yanjie Tian
  • , Tiancheng Feng
  • Northwest Institute of Nuclear Technology

科研成果: 期刊稿件文章同行评审

摘要

It is difficult to produce the radioactive standard gas source such as 133Xe, which makes the calibration of γspectrometers difficult. The uniform falling velocities of the 133Ba disk source and 137Cs point source were used to calibrate the detection efficiency of 133Xe volume source and 137Cs linear source. The non-uniform falling velocities of 133Ba disk source and 137Cs point source were used to calibrate the detection efficiency of heterogeneous 133Xe volume source and heterogeneous 137Cs linear source. The results show that the detection efficiencies of 133Xe were 0.155 (for ■46.8 mm×20 mm sample) and 0.143 (for ■63.5 mm×20 mm sample) respectively, obtained for the uniform falling velocity of the source, and the uncertainty was 8.4% (k=2). The deviation between the efficiency of the heterogeneously distributed volume source and the efficiency of uniform volume source in 14%, and 23% for linear sources. Therefore, the deviation caused by heterogeneity must be corrected.