跳到主要导航 跳到搜索 跳到主要内容

Direct-radiation response of CdZnTe detector to 14 MeV pulsed neutron beam

  • Hong Bo Pan
  • , Xiao Ping Ouyang
  • , Xian Peng Zhang
  • , Da Hai Li
  • , Hong Yun Li
  • , Meng Tong Qiu
  • , Gang Li
  • Northwest Institute of Nuclear Technology

科研成果: 期刊稿件文章同行评审

2 引用 (Scopus)

摘要

CdZnTe (CZT) crystal is one of the most attractive materials used for the new type radiation detector due to its excellent material properties. The direct-radiation response of CZT detector to 14 MeV neutron was calculated by Monte-Carlo code MCNP. The measurement of the CZT detector to 14 MeV neutron was performed on ING-103 pulsed neutron generator. The results indicate that the theoretical value accords with the experimental result. When coverage factor is 2, the expand uncertainty is less than 14%. With the sensitivity of 1.28×10-16 C·cm2 to 14 MeV neutron, the CZT detector can be applied to the high-intensity pulsed neutron measurement.

源语言英语
页(从-至)381-384
页数4
期刊Yuanzineng Kexue Jishu/Atomic Energy Science and Technology
43
4
出版状态已出版 - 4月 2009