Direct determination of Ca, Mg and Fe in serum by using a simultaneous microwave plasma torch spectrometer coupled with an improved electrothermal vaporization device

  • Jie Jiang
  • , Qiang Fei
  • , Guo Dong Feng
  • , Wen Chao Dou
  • , Ming Li
  • , Qin Han Jin
  • , Wei Jin
  • Jilin University

科研成果: 期刊稿件 › 文章 › 同行评审

4 引用 (Scopus)

摘要

A novel procedure for the direct determination of Ca, Mg and Fe in serum using an improved electrothermal vaporization simultaneous microwave plasma torch spectrometer (ETV-MPT) was described. The background emission and spectral interferences caused by tantalum ablation were reduced by coating the coil with tantalum carbide. The proposed method eliminated the need for a lengthy sample digestion procedure, and avoided the matrix interference. The operating parameters were studied and optimized. The detection limits were 0. 04, 0.4 and 4. 6 Mu;g/L for Ca, Mg and Fe, respectively, and the relative standard deviations were less than 5 % ( n = 6). The new system was successfully applied to the analysis of practical serum sample.

源语言英语
页(从-至)91-94
页数4
期刊Fenxi Huaxue/ Chinese Journal of Analytical Chemistry
卷35
期1
出版状态已出版 - 1月 2007
已对外发布是