摘要
Objective: To design a constant-force extrusion unit to meet the functional requirements of the prosthesis socket. Methods: Based on the operational principle that the constant-force derives from offsetting of the positive and negative stiffness, the leaf spring was simulated based on the finite element analysis software. According to the simulation result of the leaf spring, the coefficient of spring stiffness was calculated. The constant-force unit after processing and assembling was tested by compression experiment. Results: The simulation result was consistent with experimental result and the constant force of extrusion unit obtained an error about 2% within 3.5-mm range. Conclusions: The constant-force characteristic is confirmed by experiment, which lays the foundation of realizing the time-shared and load-bearing hand prosthesis socket.
| 源语言 | 英语 |
|---|---|
| 页(从-至) | 520-524 |
| 页数 | 5 |
| 期刊 | Yiyong Shengwu Lixue/Journal of Medical Biomechanics |
| 卷 | 31 |
| 期 | 6 |
| DOI | |
| 出版状态 | 已出版 - 1 12月 2016 |
| 已对外发布 | 是 |
学术指纹
探究 'Design of constant-force extrusion unit for hand prosthesis socket' 的科研主题。它们共同构成独一无二的指纹。引用此
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver