TY - JOUR
T1 - Design and optimization of a compact 2.45 GHz ECR ion source for BNCT applications
AU - Yang, Yang
AU - Zhou, Wenming
AU - Su, Xingkang
AU - Gu, Long
AU - Wang, Guan
AU - Zhang, Lu
AU - Yun, Di
N1 - Publisher Copyright:
© 2025 IOP Publishing Ltd and Sissa Medialab. All rights, including for text and data mining, AI training, and similar technologies, are reserved.
PY - 2025/6/1
Y1 - 2025/6/1
N2 - The 2.45 GHz electron cyclotron resonance ion source (ECRIS) is highly effective for generating high-intensity H+ ion beams. To meet the precision requirements of medical applications such as boron neutron capture therapy (BNCT), compact and adjustable ECR ion sources are essential. This study presents the design, assembly, and preliminary testing of a high-performance, compact ECR ion source. Using a three-dimensional model, we optimized the plasma chamber geometry to improve spatial efficiency and beam quality, achieving an electron density of approximately 1018 m-3 and H+ purity over 90%. Initial experiments confirmed stable and reliable operation, with a maximum extraction current of 40 mA. By integrating simulations and experimental analysis, we systematically examined the effects of working gas pressure and microwave power on ion source performance. The optimized configuration demonstrated superior stability and beam intensity compared to conventional designs. The Accelerator-Based BNCT (AB-BNCT) system employing this ion source has completed assembly and preliminary testing, showing excellent operational performance.
AB - The 2.45 GHz electron cyclotron resonance ion source (ECRIS) is highly effective for generating high-intensity H+ ion beams. To meet the precision requirements of medical applications such as boron neutron capture therapy (BNCT), compact and adjustable ECR ion sources are essential. This study presents the design, assembly, and preliminary testing of a high-performance, compact ECR ion source. Using a three-dimensional model, we optimized the plasma chamber geometry to improve spatial efficiency and beam quality, achieving an electron density of approximately 1018 m-3 and H+ purity over 90%. Initial experiments confirmed stable and reliable operation, with a maximum extraction current of 40 mA. By integrating simulations and experimental analysis, we systematically examined the effects of working gas pressure and microwave power on ion source performance. The optimized configuration demonstrated superior stability and beam intensity compared to conventional designs. The Accelerator-Based BNCT (AB-BNCT) system employing this ion source has completed assembly and preliminary testing, showing excellent operational performance.
KW - Instrumentation for neutron sources
KW - Instrumentation for particle accelerators and storage rings - low energy (linear accelerators, cyclotrons, electrostatic accelerators)
KW - Instrumentation for particle-beam therapy
KW - Ion sources (positive ions, negative ions, electron cyclotron resonance (ECR), electron beam (EBIS))
UR - https://www.scopus.com/pages/publications/105008973662
U2 - 10.1088/1748-0221/20/06/P06032
DO - 10.1088/1748-0221/20/06/P06032
M3 - 文章
AN - SCOPUS:105008973662
SN - 1748-0221
VL - 20
JO - Journal of Instrumentation
JF - Journal of Instrumentation
IS - 6
M1 - P06032
ER -