TY - JOUR
T1 - Comparison of a fixed-grid and arbitrary Lagrangian–Eulerian methods on modelling fluid–structure interaction of the aortic valve
AU - Joda, Akram
AU - Jin, Zhongmin
AU - Summers, Jon
AU - Korossis, Sotirios
N1 - Publisher Copyright:
© IMechE 2019.
PY - 2019/5/1
Y1 - 2019/5/1
N2 - This study was aimed at assessing the robustness of a fixed-grid fluid–structure interaction method (Multi-Material Arbitrary Lagrangian–Eulerian) to modelling the two-dimensional native aortic valve dynamics and comparing it to the Arbitrary Lagrangian–Eulerian method. For the fixed-grid method, the explicit finite element solver LS-DYNA was utilized, where two independent meshes for the fluid and structure were generated and the penalty method was used to handle the coupling between the fluid and structure domains. For the Arbitrary Lagrangian–Eulerian method, the implicit finite element solver ADINA was used where two separate conforming meshes were used for the valve structure and the fluid domains. The comparison demonstrated that both fluid–structure interaction methods predicted accurately the valve dynamics, fluid flow, and stress distribution, implying that fixed-grid methods can be used in situations where the Arbitrary Lagrangian–Eulerian method fails.
AB - This study was aimed at assessing the robustness of a fixed-grid fluid–structure interaction method (Multi-Material Arbitrary Lagrangian–Eulerian) to modelling the two-dimensional native aortic valve dynamics and comparing it to the Arbitrary Lagrangian–Eulerian method. For the fixed-grid method, the explicit finite element solver LS-DYNA was utilized, where two independent meshes for the fluid and structure were generated and the penalty method was used to handle the coupling between the fluid and structure domains. For the Arbitrary Lagrangian–Eulerian method, the implicit finite element solver ADINA was used where two separate conforming meshes were used for the valve structure and the fluid domains. The comparison demonstrated that both fluid–structure interaction methods predicted accurately the valve dynamics, fluid flow, and stress distribution, implying that fixed-grid methods can be used in situations where the Arbitrary Lagrangian–Eulerian method fails.
KW - Aortic valve
KW - Arbitrary Lagrangian–Eulerian
KW - finite element
KW - fixed-grid FSI
KW - fluid–structure interaction
UR - https://www.scopus.com/pages/publications/85064560654
U2 - 10.1177/0954411919837568
DO - 10.1177/0954411919837568
M3 - 文章
C2 - 30922162
AN - SCOPUS:85064560654
SN - 0954-4119
VL - 233
SP - 544
EP - 553
JO - Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
JF - Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
IS - 5
ER -