跳到主要导航 跳到搜索 跳到主要内容

Calibration of cone beam rotational X-ray image sequence

  • Hangzhou Dianzi University
  • Xi'an Jiaotong University

科研成果: 期刊稿件文章同行评审

16 引用 (Scopus)

摘要

The real X-ray projection does not abide by Lambert-Beer Law, since the X-ray is polychromatic and the imaging chains are nonlinear. Based on the generating process of X-ray images, an equivalent nonlinear transform model is firstly proposed which considers all the nonlinear factors as one nonlinear transform. Then the 3D (three-dimensional) X-ray projection of cone beam is defined. The constraints of Radon transform, named H-L (Helgasson-ludwig) consistency conditions, are expanded to fan-beam. After that an algorithm is developed to calibrate Rotational X-ray image sequence (RXIS). The algorithm uses a set of exponential functions to approximate the nonlinear inverse transform. According to expanded H-L consistency conditions, finally a kind of nonlinear measure for RXIS is defined. Experimental results show that the proposed algorithm can decrease the nonlinear measure to below 0.01.

源语言英语
页(从-至)424-428
页数5
期刊Chinese Journal of Electronics
13
3
出版状态已出版 - 7月 2004

学术指纹

探究 'Calibration of cone beam rotational X-ray image sequence' 的科研主题。它们共同构成独一无二的学术指纹。

引用此