摘要
This paper presents an optical readout method based uncooled infrared imaging system that contains an optical read-out section and a bi-material micro-cantilever arrays (BMCA) detector. The optical read-out section is based on incoherent light spatial filter technique. The read-out section converts the deflection angles of BMCA caused by absorption of infrared (IR) radiation to a visible image through optical filtering operation of the BMCA with a knife-edge filter. A spatial mathematical model is presented to describe the read-out method and its validity is proved. The IR image of a person's hand obtained through the using of the 100∈×∈100 BMCA and the 12-bit A/D quantizer demonstrates the ability of the system to create image. The performance test shows that the average Noise-equivalent temperature difference (NETD) of the imaging system can arrive at about 183mK with some areas having a NETD as low as 78mK.
| 源语言 | 英语 |
|---|---|
| 页(从-至) | 261-271 |
| 页数 | 11 |
| 期刊 | International Journal of Infrared and Millimeter Waves |
| 卷 | 29 |
| 期 | 3 |
| DOI | |
| 出版状态 | 已出版 - 3月 2008 |
| 已对外发布 | 是 |
学术指纹
探究 'An optical readout method based uncooled infrared imaging system' 的科研主题。它们共同构成独一无二的学术指纹。引用此
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver