An energy spectrometer for detecting flash X-ray spectra from 200 to 3000keV

  • Zhao Feng Su
  • , Hai Liang Yang
  • , Ai Ci Qiu
  • , Jian Feng Sun
  • , Pei Tian Cong
  • , Liang Ping Wang
  • , Tian Shi Lei
  • , Juan Juan Han
  • Northwest Institute of Nuclear Technology

科研成果: 会议稿件论文同行评审

摘要

The basic theory and experimental results obtained in measuring energy spectra of time- resolved high energy pulsed X-ray are reported. MCNP code was employed to calculate the energy deposition of high energy photons on the energy spectrometer. The waveforms of high energy pulsed X-ray created on the "Qiangguang-I" accelerator were measured by using the energy spectrometer. According to the deposited energy and integral data, the time-resolved high pulsed x-ray energy spectrum was obtained. By using the voltage and current waveforms of the diode, the energy spectrum was calculated in details. The results were agreed with that measured with the experimental energy spectrum.

源语言英语
出版状态已出版 - 2008
已对外发布
活动17th International Conference on High Power Particle Beams, BEAMS'08 - Xi'an, 中国
期限: 6 7月 200811 7月 2008

会议

会议17th International Conference on High Power Particle Beams, BEAMS'08
国家/地区中国
Xi'an
时期6/07/0811/07/08