基于虚拟点探测器的虚拟点源效率刻度技术

  • Zining Tian
  • , Wenbiao Liu
  • , Wei Chen
  • , Bin Long
  • , Tiancheng Feng
  • , Xuemei Wang
  • , Xiaoping Ouyang
  • Northwest Institute of Nuclear Technology

科研成果: 期刊稿件文章同行评审

2 引用 (Scopus)

摘要

The detection efficiency of 75 mm×25 mm of 241Am standard samples were studied by using laboratory HPGe detection. When we used detectors with the same type of structure to measure a given volume source, the linearity function between the virtual point source and the virtual point detector was established by using LabSOCS software and Monte Carlo method. For a detector, the position of virtual point detector can be fixed by a point source placed at different height, and then, the position of virtual point source can be fixed by using their linearity function. The real calibration process will be carried out by placing a point source at its virtual position, namely, the volume source is replaced with the point source. In conclusion, the position of virtual point source can be obtained through the virtual point detector. This is a completely new technology method, which will be widely applied in radioactive measurement.

投稿的翻译标题Virtual point source efficiency calibration technology based on virtual point detector
源语言繁体中文
期刊论文编号034001
期刊Qiangjiguang Yu Lizishu/High Power Laser and Particle Beams
30
3
DOI
出版状态已出版 - 1 3月 2018
已对外发布

关键词

  • Efficiency calibration
  • Monte Carlo method
  • Virtual point detector
  • Virtual point source