Value of a novel Y-Z magnetic totally implantable venous access port in improving the success rate of one-time needle insertion

  • Miaomiao Zhang
  • , Xin Lyu
  • , Juanjuan Wang
  • , Aihua Shi
  • , Yunyun Zheng
  • , Yi Lyu
  • , Xiaopeng Yan

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

Background and objectives: A totally implantable venous access port (TIVAP) is a commonly used intravenous infusion device for patients receiving chemotherapy or long-term infusion therapy. To improve the success rate of one-time insertion of the Huber needle, we developed a novel Y-Z magnetic TIVAP (Y-Z MTIVAP), which we produced using three-dimensional printing technology. Materials and methods: The Y-Z MTIVAP includes a magnetic port body and a magnetic positioning device. For testing, we established four venous port implantation models using the two types of TIVAPs and two implantation depth ranges (≤5 mm and >5 mm). Twenty nurses performed Huber needle puncture with the four models, and we recorded the number of attempts required for successful needle insertion, the operation time, and the operator's satisfaction. Results: The success rate for one-time needle insertion with the Y-Z MTIVAP was significantly higher than that with the traditional TIVAP at either depth range (100% vs. 75% at ≤5 mm, p = 0.047; 95% vs. 35% at >5 mm, p < 0.001). With increasing implantation depth, the success rate for one-time insertion was significantly reduced with the traditional TIVAP (75% at ≤5 mm vs. 35% vs. >5 mm, p = 0.025), but the success rate with the Y-Z MTIVAP was not significantly affected (100% vs. 95%, p = 1.000). The operation time with the Y-Z MTIVAP was significantly shorter than that with the traditional TIVAP at either depth range (both p < 0.001), and 90% of operators reported that the Y-Z MTIVAP was superior to the traditional TIVAP. Conclusions: The theoretical design of Y-Z MTIVAP is feasible, and the preliminary in vitro simulation experiment shows that it can significantly improve puncture success rate and shortened operation time.

Original languageEnglish
Article number1241780
JournalFrontiers in Surgery
Volume10
DOIs
StatePublished - 2023
Externally publishedYes

Keywords

  • Huber needle
  • magnet
  • magnetic anchor technique
  • magnetosurgery
  • total implantable venous access port

Fingerprint

Dive into the research topics of 'Value of a novel Y-Z magnetic totally implantable venous access port in improving the success rate of one-time needle insertion'. Together they form a unique fingerprint.

Cite this