Systematic analysis of the quench process performance and simulation of the 9.4 T-800 mm whole-body MRI magnet

  • Shunzhong Chen
  • , Lei Wang
  • , Zili Zhang
  • , Yaohui Wang
  • , Hui Wang
  • , Junsheng Cheng
  • , Yinming Dai
  • , Qiuliang Wang

Research output: Contribution to journalArticlepeer-review

3 Scopus citations

Abstract

In 2021, the Institute of Electrical Engineering, Chinese Academy of Sciences successfully reached 9.4 T in a whole-body magnetic resonance imaging (MRI) superconducting magnet with an inner diameter of 800 mm. In this study, a systematic analysis of both the real quench protection performance and a simulation are reported. The four successful quench protections during the entire energization process proved the feasibility of the ‘in-out-in’ quench protection protocol for a 9.4 T-800 mm superconducting magnet. The quench trigger sequence was shown to be adjusted by changing the heater thickness, which demonstrates the flexibility of the ‘in-out-in’ quench protection protocol to fit a different MRI magnet design. The high accuracy of the quench protection simulation method and code was confirmed through comparison of the simulation results with the real performance. Moreover, the limitations of the current quench protection and reference value to other MRI magnets were discussed. It is believed that this study will be useful to other research groups and promote the development of an extremely high-field whole-body MRI system.

Original languageEnglish
Article number045007
JournalSuperconductor Science and Technology
Volume36
Issue number4
DOIs
StatePublished - Apr 2023

Keywords

  • 9.4 T-800 mm
  • quench protection
  • whole-body MRI

Fingerprint

Dive into the research topics of 'Systematic analysis of the quench process performance and simulation of the 9.4 T-800 mm whole-body MRI magnet'. Together they form a unique fingerprint.

Cite this