Abstract
Abstract Accurate control and measurement of real-time sample temperature are critical for the understanding and interpretation of the experimental results from in situ heating experiments inside environmental transmission electron microscope (ETEM). However, quantifying the real-time sample temperature remains a challenging task for commercial in situ TEM heating devices, especially under gas conditions. In this work, we developed a home-made micro-electrical-mechanical-system (MEMS) heater with unprecedented small temperature gradient and thermal drift, which not only enables the temperature evolution caused by gas injection to be measured in real-time but also makes the key heat dissipation path easier to model to theoretically understand and predict the temperature decrease. A new parameter termed as gas cooling ability (H), determined purely by the physical properties of the gas, can be used to compare and predict the gas-induced temperature decrease by different gases. Our findings can act as a reference for predicting the real temperature for in situ heating experiments without closed-loop temperature sensing capabilities in the gas environment, as well as all gas-related heating systems.
| Original language | English |
|---|---|
| Pages (from-to) | 758-766 |
| Number of pages | 9 |
| Journal | Microscopy and Microanalysis |
| Volume | 27 |
| Issue number | 4 |
| DOIs | |
| State | Published - Aug 2021 |
Keywords
- Key words environmental TEM (ETEM)
- MEMS heater
- gas cooling effect
- in situ heating
- temperature measurement
Fingerprint
Dive into the research topics of 'Quantifying Real-Time Sample Temperature under the Gas Environment in the Transmission Electron Microscope Using a Novel MEMS Heater'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver