Model-based tungsten anode X-ray tube spectra reconstruction method

Research output: Contribution to journalArticlepeer-review

6 Scopus citations

Abstract

Knowledge of X-ray spectra is essential for X-ray dose control, X-ray image quality evaluation and dual-energy X-ray imaging etc. A model-based tungsten anode X-ray tube spectra reconstruction method is presented in this paper. A physical meaningful seven-parameter X-ray tube spectra model is firstly established. Then model parameters are estimated from attenuation data so that X-ray tube spectra can be reconstructed. In the experiment, a series of aluminum plates and copper plates with different thickness are selected as attenuator for a selected tube voltage. Model parameters are estimated from experimentally measured attenuation data to reconstruct X-ray tube spectra. Experiments show that the method is easy to apply and can reconstruct both bremsstrahlung and characteristic spectra with only a few attenuation data. The difference of attenuation data between initial spectra and reconstructed spectra is lower than 0.3%. The difference between modified handbook spectra and reconstructed spectra is lower than 5%.

Original languageEnglish
Pages (from-to)2285-2291
Number of pages7
JournalTien Tzu Hsueh Pao/Acta Electronica Sinica
Volume38
Issue number10
StatePublished - Oct 2010

Keywords

  • Attenuation data
  • Spectra reconstruction
  • X-ray tube spectra model

Fingerprint

Dive into the research topics of 'Model-based tungsten anode X-ray tube spectra reconstruction method'. Together they form a unique fingerprint.

Cite this