LIBS測定におけるスペクトル強度の照射ごとのばらつきの誤差伝播による解析

Translated title of the contribution: Analysis of Pulse-to-Pulse Signal Uncertainty in LIBS Spectral Intensity on the Basis of Error Propagation

Research output: Contribution to journalReview articlepeer-review

Abstract

Atomic spectral lines obtained by laser-induced breakdown spectroscopy (LIBS) often suffer from serious pulse-to-pulse fluctuation, which limits the accuracy of the quantitative analysis. Solving this problem is an important issue for improving the analytical performance of LIBS. In the present review the model to simulate the emission spectral intensity of LIBS measurements is introduced, and the method to evaluate the propagation of the pulse-to-pulse variation of plasma parameters to the variation of emission spectral intensity by the error propagation analysis is explained. The recent three studies found in the literature that investigated the pulse-to-pulse fluctuation of the spectral line intensities on the basis of error propagation analysis are reviewed.

Translated title of the contributionAnalysis of Pulse-to-Pulse Signal Uncertainty in LIBS Spectral Intensity on the Basis of Error Propagation
Original languageJapanese
Pages (from-to)35-40
Number of pages6
JournalTetsu-To-Hagane/Journal of the Iron and Steel Institute of Japan
Volume110
Issue number2
DOIs
StatePublished - 2024
Externally publishedYes

Fingerprint

Dive into the research topics of 'Analysis of Pulse-to-Pulse Signal Uncertainty in LIBS Spectral Intensity on the Basis of Error Propagation'. Together they form a unique fingerprint.

Cite this