Influence of projection data correction on digital breast tomosynthesis imaging

  • Xin Yu Zhang
  • , Hua Zhang
  • , Zhao Ying Bian
  • , Dong Zeng
  • , Ji He
  • , Xiu Mei Tian
  • , Jian Hua Ma
  • , Jing Huang
  • Southern Medical University

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

OBJECTIVE: To investigate the effect of detector performance during digital breast tomography (DBT) projection data acquisition on reconstructed image quality. METHODS: With reference to the traditional detector data correction method and the specific data acquisition pattern in DBT imaging, we utilized dark field correction, light field and its gain correction for processing the projection data collected by the detector. The reconstructed images were evaluated using iterative reconstruction method based on total generalized variation (TGV). RESULTS: In physical breast phantom experiment, the proposed method resulted in a reduced Heel effect caused by nonuniform photon number. The reconstructed DBT images after correction showed obviously improved image quality especially in the details with a low contrast. CONCLUSION: The dark field correction, light field and its gain correction process for DBT image reconstruction can improve the image quality.

Original languageEnglish
Pages (from-to)323-329
Number of pages7
JournalNan Fang Yi Ke Da Xue Xue Bao / Journal of Southern Medical University
Volume37
Issue number3
StatePublished - 20 Mar 2017

Fingerprint

Dive into the research topics of 'Influence of projection data correction on digital breast tomosynthesis imaging'. Together they form a unique fingerprint.

Cite this