Abstract
OBJECTIVE: We propose a method using total variation (TV) regularization in deconvolution for partial volume correction in PET imaging. In the degraded image model, we used TV regularization procedure in Van Cittert (VC) and Richardson-Lucy (RL) deconvolution algorithms. These methods were tested in simulated NCAT images and images of NEMA NU4-2008 IQ phantom and tumor-bearing mouse scanned by Simens Invoen microPET. The simulated experiment and tumor-bearing mouse experiment showed that the algorithms using TV regularization provided superior qualitative and quantitative appearance compared with traditional VC and RL algorithms. When the mean intensity of the tumor increased by (10±1.8)%, the SD increase percentage was decreased from 49.98% to 14.26% and from 42.76% to 4.70%, suggesting the efficiency of the proposed algorithms for reducing PVEs in PET.
| Original language | English |
|---|---|
| Pages (from-to) | 474-480 |
| Number of pages | 7 |
| Journal | Nan Fang Yi Ke Da Xue Xue Bao / Journal of Southern Medical University |
| Volume | 35 |
| Issue number | 4 |
| State | Published - 1 Apr 2015 |
| Externally published | Yes |
Fingerprint
Dive into the research topics of 'Incorporation of TV regularization in deconvolution for partial volume correction in PET imaging'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver