TY - JOUR
T1 - Evaluation of dentinal tubule occlusion and pulp tissue response after using 980-nm diode laser for dentin hypersensitivity treatment
AU - Meng, Yuchen
AU - Huang, Fan
AU - Wang, Silin
AU - Huang, Xin
AU - Lu, Yi
AU - Li, Yuncong
AU - Dong, Yulin
AU - Pei, Dandan
N1 - Publisher Copyright:
© 2023, The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature.
PY - 2023/8
Y1 - 2023/8
N2 - Objectives: To evaluate the effectiveness of the 980-nm diode laser for dentinal tubule occlusion, measure the intrapulpal temperature, and investigate the dental pulp response. Materials and methods: The dentinal samples were randomly divided into G1–G7 groups: control; 980-nm laser irradiation (0.5 W, 10 s; 0.5 W, 10 s × 2; 0.8 W, 10 s; 0.8 W, 10 s × 2; 1.0 W, 10 s; 1.0 W, 10 s × 2). The dentin discs were applied for laser irradiation and analyzed by scanning electron microscopy (SEM). The intrapulpal temperature was measured on the 1.0-mm and 2.0-mm thickness samples, and then divided into G2–G7 groups according to laser irradiation. Moreover, forty Sprague Dawley rats were randomly divided into the laser-irradiated group (euthanized at 1, 7, and 14 days after irradiation) and the control group (non-irradiated). qRT-PCR, histomorphology, and immunohistochemistry analysis were employed to evaluate the response of dental pulp. Results: SEM indicated the occluding ratio of dentinal tubules in the G5 (0.8 W, 10 s × 2) and G7 (1.0 W, 10 s × 2) were significantly higher than the other groups (p < 0.05). The maximum intrapulpal temperature rises in the G5 were lower than the standard line (5.5 ℃). qRT-PCR showed that the mRNA expression level of TNF-α and HSP-70 upregulated significantly at 1 day (p < 0.05). Histomorphology and immunohistochemistry analysis showed that, compared with the control group, the inflammatory reaction was slightly higher at the 1 and 7 days (p < 0.05) and decreased to the normal levels at 14 days (p > 0.05). Conclusions: A 980-nm laser at a power of 0.8 W with 10 s × 2 defines the best treatment for dentin hypersensitivity in terms of compromise between the efficacy of the treatment and the safety of the pulp. Clinical relevance: The 980-nm laser is an effective option for treating dentin sensitivity. However, we need to ensure the safety of the pulp during laser irradiation.
AB - Objectives: To evaluate the effectiveness of the 980-nm diode laser for dentinal tubule occlusion, measure the intrapulpal temperature, and investigate the dental pulp response. Materials and methods: The dentinal samples were randomly divided into G1–G7 groups: control; 980-nm laser irradiation (0.5 W, 10 s; 0.5 W, 10 s × 2; 0.8 W, 10 s; 0.8 W, 10 s × 2; 1.0 W, 10 s; 1.0 W, 10 s × 2). The dentin discs were applied for laser irradiation and analyzed by scanning electron microscopy (SEM). The intrapulpal temperature was measured on the 1.0-mm and 2.0-mm thickness samples, and then divided into G2–G7 groups according to laser irradiation. Moreover, forty Sprague Dawley rats were randomly divided into the laser-irradiated group (euthanized at 1, 7, and 14 days after irradiation) and the control group (non-irradiated). qRT-PCR, histomorphology, and immunohistochemistry analysis were employed to evaluate the response of dental pulp. Results: SEM indicated the occluding ratio of dentinal tubules in the G5 (0.8 W, 10 s × 2) and G7 (1.0 W, 10 s × 2) were significantly higher than the other groups (p < 0.05). The maximum intrapulpal temperature rises in the G5 were lower than the standard line (5.5 ℃). qRT-PCR showed that the mRNA expression level of TNF-α and HSP-70 upregulated significantly at 1 day (p < 0.05). Histomorphology and immunohistochemistry analysis showed that, compared with the control group, the inflammatory reaction was slightly higher at the 1 and 7 days (p < 0.05) and decreased to the normal levels at 14 days (p > 0.05). Conclusions: A 980-nm laser at a power of 0.8 W with 10 s × 2 defines the best treatment for dentin hypersensitivity in terms of compromise between the efficacy of the treatment and the safety of the pulp. Clinical relevance: The 980-nm laser is an effective option for treating dentin sensitivity. However, we need to ensure the safety of the pulp during laser irradiation.
KW - 980-nm diode laser
KW - Dentin hypersensitivity
KW - Dentinal tubule occlusion
KW - Intrapulpal temperature
KW - Pulp response
UR - https://www.scopus.com/pages/publications/85163691660
U2 - 10.1007/s00784-023-05114-y
DO - 10.1007/s00784-023-05114-y
M3 - 文章
C2 - 37382717
AN - SCOPUS:85163691660
SN - 1432-6981
VL - 27
SP - 4843
EP - 4854
JO - Clinical Oral Investigations
JF - Clinical Oral Investigations
IS - 8
ER -