Abstract
A new definition of the experimental wear factor was established and reported as a function of cross-shear motion and contact pressure using a multi-directional pin-on-plate wear testing machine for conventional polyethylene in the present study. An independent computational wear model was developed by incorporating the cross-shear motion and contact pressure-dependent wear factor into the Archard's law, in additional to load and sliding distance. The computational prediction of wear volume was directly compared with a simulator testing of a polyethylene hip joint with a 28 mm diameter. The effect of increasing the femoral head size was subsequently considered and was shown to increase wear, as a result of increased sliding distance and reduced contact pressure.
| Original language | English |
|---|---|
| Pages (from-to) | 912-918 |
| Number of pages | 7 |
| Journal | Journal of Biomechanics |
| Volume | 42 |
| Issue number | 7 |
| DOIs | |
| State | Published - 11 May 2009 |
| Externally published | Yes |
Keywords
- Contact pressure
- Cross-shear
- Head diameter
- UHMWPE hip
- Wear modelling
Fingerprint
Dive into the research topics of 'Enhanced computational prediction of polyethylene wear in hip joints by incorporating cross-shear and contact pressure in additional to load and sliding distance: Effect of head diameter'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver